This equipment is used for performance testing of batteries that have undergone formation, to obtain their capacity, internal resistance, and other key parameters. It also classifies and grades the batteries.
Indicator | Parameter |
---|---|
Channel Number | 8 |
Input Power | AC 220V ±10% / 50Hz |
Input Active Power | 400W |
Resolution | AD: 24-bit; DA: 16-bit |
Voltage | Constant voltage range control |
Voltage Range | 0.025V~5V |
Minimum Discharge Voltage | Both ends of the upper and lower fixtures can discharge to 1V; 1.5V with 2m cable length |
Accuracy | ± 0.05% of FS |
Stability | ± 0.05% of FS |
Current | Output range per channel |
Range 1 | 0.5mA-0.1A |
Range 2 | 0.1A-3A |
Range 3 | 3A-6A |
Accuracy | ± 0.1% of FS |
Constant Voltage Cutoff Current | |
Range 1 | 0.2mA |
Range 2 | 6mA |
Range 3 | 12mA |
Stability | ± 0.05% of FS |
Power | Maximum output power per channel |
Stability | ± 0.1% of FS |
Time | Current rise time |
Step time range | ≤(365*24) hours/step |
Time format | Support 00:00:00:00 (h, min, s, ms) |
Data Recording | |
Recording conditions | Minimum time interval: 100ms |
Minimum voltage interval: 10mV | |
Minimum current interval | |
Range 1 | 0.2mA |
Range 2 | 6mA |
Range 3 | 12mA |
Recording frequency | 10Hz |
Charging | |
Charging modes | Constant current charging, constant voltage charging, constant current constant voltage charging, constant power charging |
Cutoff conditions | Voltage, current, relative time, capacity, energy, -△V |
Discharging | |
Discharging modes | Constant current discharge, constant voltage discharge, constant current constant voltage discharge, constant power discharge, constant resistance discharge |
Cutoff conditions | Voltage, current, relative time, capacity, energy |
Pulse Mode | |
Charging | Constant current mode, constant power mode |
Discharging | Constant current mode, constant power mode |
Minimum pulse width | 500ms |
Number of pulses | Each pulse step supports 32 different pulses |
Continuous charge-discharge switching | A pulse step can achieve continuous switching from charging to discharging |
Cutoff conditions | Voltage, relative time |
DCIR Test | Supports custom points for DCIR calculation |
Cycle | |
Cycle test range | 1~65535 times |
Number of steps per cycle | 254 |
Nested cycles | Supports nested cycles, up to 3 layers |
Protection | |
Power-off data protection | |
Offline test function | |
Safety protection conditions | Settable parameters include: upper voltage limit, lower voltage limit, upper current limit, lower current limit, upper capacity limit, delay time, reverse polarity protection |
IP Protection Level | IP20 |
Channel Features | Dual closed-loop structure for constant current and constant voltage sources |
Channel Control Mode | Independent control |
Voltage and Current Detection Sampling | Four-wire connection |
Noise | <55dB |
Database | MySQL database for centralized test data management |
Upper Computer Communication Mode | Based on TCP/IP protocol |
Server Operating System | Windows 7/Windows 10 |
Data Output Format | EXCEL2003, 2010, TXT |
Server Disk Configuration | 500GB |
Communication Interface | Network port |
Leakage Current | <5μA |
Power Supply: AC220V/50Hz, power approx. 400W.
Equipment Specifications:
Equipment Weight: approx. 15kg.
Single Machine Dimensions: approx. L (490mm) x W (400mm) x H (130mm).
Appearance Color: Light Gray.
Environmental Requirements:
Ambient Temperature: 25±5℃.
Relative Humidity: 30%~75% RCH, ensure the site is dry and well-ventilated, free from acid and alkali corrosion.
Grounding: Protective grounding.